Tyco Kendall Genius 2 Operation And Service Manual

Type
Operation And Service Manual
Inhoudsopgave
Pagina
Deel I - Functiebeschrijving ............................................................................................................................................................. 1
Deel II — Veiligheid en waarschuwingen ..................................................................................................................................... 2
Sectie III - Pictogrammen en hun betekenis ................................................................................................................................. 3
Deel IV - Vereist materiaal ............................................................................................................................................................... 4
Deel V — Systeeminstallatie .......................................................................................................................................................... 4
Deel VI - Gebruiksprocedure ............................................................................................................................................................ 5
Stroom inschakelen ....................................................................................................................................................................... 5
Opstartscherm ............................................................................................................................................................................... 5
Herinneringscherm ....................................................................................................................................................................... 5
Hoofdmenu ................................................................................................................................................................................... 5
Voorkeuren .................................................................................................................................................................................... 5
Taalkeuze ...................................................................................................................................................................................... 5
Gebruiker/locatie-informatie instellen ......................................................................................................................................... 6
Tijd en datum instellen .................................................................................................................................................................... 6
Kalibratie controleren ...................................................................................................................................................................... 6
Deel VII - Foutmeldingen ................................................................................................................................................................. 9
Deel VIII - Specificaties van controle/kalibrator ......................................................................................................................... 10
Deel IX - Reinigingsinstructies ...................................................................................................................................................... 10
Sectie X - Probleemoplossing ........................................................................................................................................................ 11
Sectie XI - Klantenservice ............................................................................................................................................................... 12
Sectie XII — Garantie ..................................................................................................................................................................... 12
Sectie XIII — Verklaring elektromagnetische conformiteit .................................................................................................... 13
Dit product bevat software waarvan Tyco Healthcare Group LP de enige eigenaar is. Tyco Healthcare Group LP verleent de gebruiker een
niet-exclusieve en beperkte licentie om de software te gebruiken in overeenstemming met de bedieningsinstructies. Een kopie van de
licentie kan worden verkregen bij Tyco Healthcare Group LP.
Nederlands
Status: Approved Doc#: AW10001726 Effective Date: nulldate
10/23/2008 Artwork#s: WN 182563-01 Job#: 16245
1
GENIUS
2 Controle/kalibrator
Gebruiks- en onderhoudshandleiding
Deel I - Functiebeschrijving
Deze handleiding beschrijft de werking van de GENIUS 2-controle/kalibrator. Dit apparaat is ontwikkeld om de kalibratie van de
GENIUS 2 tympanische thermometer te controleren en indien nodig automatisch de thermometer opnieuw te kalibreren. Daarnaast
beschikt de GENIUS 2-controle/kalibrator over de mogelijkheid om testrapporten te genereren voor geteste thermometers en deze
op een USB-ashstation op te slaan. De GENIUS 2-controle/kalibrator kan gebruikt worden voor alle softwarerevisies van de GENIUS
2-thermometer.
De GENIUS 2-controle/kalibrator bevat twee onafhankelijk gecontroleerde infrarode kalibratiedoelen die lijken op de
kalibratiedoelen in de fabriek. Deze “blackbody”-doelen zijn speciaal ontworpen voor een eciënte overdracht van stralingswarmte
en geven infrarode straling af die het menselijke oor benadert bij temperaturen van 32,2 °C (90,0 °F) en 40,6 °C (105,0 °F).
GENIUS
F/C˚
F/C˚
1 2 3 4
8 910
5
6
7
11
Checker / Calibrator
F 3.15A 250V
Afbeelding 1 - Layout van voorpaneel
Legende
1 Voedingaansluiting 7 Hoge temperatuurdoel
2 Stroomschakelaar 8 LED van hoge temperatuurdoel
3 Zekeringhouder 9 LCD-display
4 LED van lage temperatuurdoel 10 Codeerknop
5 Lage temperatuurdoel 11 USB-poort
6 Connector van GENIUS 2 thermometer
Status: Approved Doc#: AW10001726 Effective Date: nulldate
10/23/2008 Artwork#s: WN 182563-01 Job#: 16245
2
GENIUS
2 Controle/kalibrator
Gebruiks- en onderhoudshandleiding
Deel II - Veiligheid en waarschuwingen
De GENIUS 2-controle/kalibrator is bedoeld voor gebruik zoals wordt aangegeven in deze handleiding. De bescherming door de
veiligheidsmaterialen in de GENIUS 2-controle/kalibrator kunnen teniet worden gedaan als het apparaat niet wordt gebruikt
volgens de aanwijzingen in deze handleiding.
De behuizing van de GENIUS 2-controle/kalibrator is niet waterdicht. Vermijd het morsen van enige vloeistof op het voorpaneel
van het apparaat. De GENIUS 2-controle/kalibrator wordt gevoed via het elektriciteitsnet. Vermijd een mogelijk contact tussen
vloeistof en enig buitenoppervlak van het apparaat wanneer deze is aangesloten op het net.
De oppervlaktekwaliteit van de binnenzijde van de referentie blackbody-doelen is zeer belangrijk voor de werking als
doeltreffende infrarode radiatoren. Men moet voorzichtig zijn bij het werken in de nabijheid van de blackbody-doelen, om zeker
te stellen dat er geen contact wordt gemaakt met het oppervlak binnen de rubberen flappen.
In het geval van een gesprongen zekering dient u de zekering te vervangen met een zekering die precies overeenkomt met de
nominale spanning, stroom en vormfactor, zoals aangegeven in het gedeelte Specificaties van GENIUS 2-controle/kalibrator van
deze handleiding. Controleer of het apparaat is losgekoppeld van het elektriciteitsnet voordat een gesprongen zekering wordt
vervangen.
Het apparaat mag niet worden gebruikt in de aanwezigheid van brandbare anesthetica. Er bestaat explosiegevaar in de nabijheid
van deze anesthetica.
Dit apparaat is niet voorzien van onderdelen die door de gebruiker kunnen worden onderhouden. Het voorpaneel niet
verwijderen. Laat het onderhoud uitvoeren door gekwalificeerd onderhoudspersoneel.
Gebruik alleen de geleverde stroommodule bij dit apparaat.
Wanneer u wordt gevraagd om de GENIUS 2-thermometer in een van de blackbody-doelen te plaatsen, heeft u 60 seconden de
tijd om deze handeling uit te voeren. De procedure zal worden afgebroken als deze periode wordt overschreden.
Zorg ervoor dat de thermometerlens schoon is voordat u een GENIUS 2-thermometer controleert en/of kalibreert. Wanneer deze
effectief is gereinigd, bent u ervan verzekerd dat de toegestane kalibratieafwijking correct wordt gecontroleerd. Raadpleeg de
reinigingsinstructies in de gebruikshandleiding van de Genius 2-thermometer.
Bij het installeren van een GENIUS 2-sondekapje op de punt van de GENIUS 2-thermometer, dient u alleen kapjes te gebruiken
die in een cassette zitten en die niet los liggen. Probeer elk handcontact met de sondekapjes te vermijden, om onnauwkeurige
aflezingen te voorkomen. Gebruik de sondekapjes niet opnieuw. Gebruik alleen GENIUS 2-sondekapjes.
Laat de GENIUS 2-kalibrator na het inschakelen gedurende ten minste 15 minuten opwarmen voor gebruik. Controleer
of de GENIUS 2-thermometers, sondekapjes en de GENIUS 2-controle/kalibrator voldoende tijd hebben gehad om op
kamertemperatuur te komen voordat u verder gaat. Zorg voor een langere opwarmperiode als het apparaat of de thermometers
een temperatuur hebben op de uiterste grenzen van de toegestane omgevingstemperatuur. De draagbaarheid van de
GENIUS 2-controle/kalibrator staat verplaatsing toe van het apparaat naar ruimten die voldoen aan de vereisten voor de
omgevingstemperatuur.
Gebruik een USB-flashstation die is voorzien van het CE-keurmerk en die voldoet aan de elektromagnetische stralingsnorm
EN55022.
Plaats de GENIUS 2-thermometer niet op het oppervlak van de GENIUS 2-controle/kalibrator. De temperatuursensoren in de
sondetip van de thermometer zullen de warmte van de “blackbody”-doelen waarnemen via het oppervlak van het voorpaneel. Dit
zorgt voor een instabiliteit van de thermometersensoren, wat zal leiden tot onjuiste temperatuuraflezingen.
De GENIUS 2-controle/kalibrator mag alleen binnenshuis worden gebruikt, in een ruimte zonder tocht en zonder grote
temperatuurswisselingen.
De USB-poort op het voorpaneel is alleen bedoeld voor gebruik met een USB-flashstation. Er mogen geen andere USB-apparaten
op de USB-poort worden aangesloten. De USB-flashstation moet voor het inschakelen van het apparaat zijn geïnstalleerd.
Retourneer de GENIUS 2-controle/kalibrator aan de fabriek voor herkalibratie indien het apparaat is gevallen
Status: Approved Doc#: AW10001726 Effective Date: nulldate
10/23/2008 Artwork#s: WN 182563-01 Job#: 16245
3
GENIUS
2 Controle/kalibrator
Gebruiks- en onderhoudshandleiding
Sectie III - Pictogrammen en hun betekenis
SN
Serienummer
CE MARK SYMBOLS
NON-CE MARK SYMBOLS
ELECTRONIC SYMBOLS AND CERTIFIED TESTING MARKS
Corrugated
Recycles
STERILE
Keep Away
From Sunlight
Keep Dry
Single Use
LOT
0123
FRAGILE
THIS SIDE UP
DEHP FREE
40°
Avoid Temperature
Extremes
Do not use if package is
opened or damaged.
STERILE R
A
STERILE
EOSTERILE
STERILE
STERILE R
A
STERILE
EOSTERILE
Humidity Limitation
Atmospheric pressure limitation
Use By
C US
LATEX FREE
Caution, consult
accompanying
documents
Consult
Instructions
For Use
MR
MR UNSAFE
NON-PYROGENIC
PROTECT FROM HEAT AND
RADIOACTIVE SOURCES
NON
STERILE
2
STERILIZE
DO NOT RESTERILIZE
POWDER FREE
POWDER
Fabrikant
Zie bijgaande
documenten
CE MARK SYMBOLS
NON-CE MARK SYMBOLS
ELECTRONIC SYMBOLS AND CERTIFIED TESTING MARKS
Corrugated
Recycles
STERILE
Keep Away
From Sunlight
Keep Dry
Single Use
LOT
0123
FRAGILE
THIS SIDE UP
DEHP FREE
40°
Avoid Temperature
Extremes
Do not use if package is
opened or damaged.
STERILE R
A
STERILE
EOSTERILE
STERILE
STERILE R
A
STERILE
EOSTERILE
Humidity Limitation
Atmospheric pressure limitation
Use By
C US
LATEX FREE
Caution, consult
accompanying
documents
Consult
Instructions
For Use
MR
MR UNSAFE
NON-PYROGENIC
PROTECT FROM HEAT AND
RADIOACTIVE SOURCES
NON
STERILE
2
STERILIZE
DO NOT RESTERILIZE
POWDER FREE
POWDER
Fabricagedatum
Stand-bystroom
Dubbele of versterkte
isolatiebescherming
Vervangen door
gelijksoortige zekering
CE-markering
USB-poort Weggooien in
overeenstemming met
richtlijnen voor elektrisch
en elektronisch afval
F/C˚
Hoge temperatuurdoel
LISTED
(LABORATORY EQUIPMENT)
46EA
UL-opname-informatie
F/C˚
Lage temperatuurdoel
Let op: Alleen voor
gebruik binnenshuis
Aansluiting van GENIUS
2-thermometer
De GENIUS 2 infrarode tympanische thermometer is een betrouwbaar en nauwkeurig apparaat voor het opnemen van de
temperatuur. De GENIUS 2-thermometer dankt zijn nauwkeurigheid aan het ontwerp, de gecontroleerde kalibratieprocedures en de
strikte fabricagemethoden. Een van de belangrijkste functionele onderdelen van een thermometer is het ACCUSYSTEM√-sondekapje
geproduceerd door Tyco Healthcare/Kendall. Het ACCUSYSTEM√-sondekapje voor de GENIUS 2-thermometer fungeert, wanneer
geplaatst op een GENIUS 2-thermometer, als een infectiepreventiedrempel tussen de patiënt en het apparaat en als medium voor
warmtetransmissie van de patiënt naar de thermometer. De functionaliteit van het ACCUSYSTEM√-sondekapje voor de
GENIUS 2-thermometer is zeer belangrijk voor het voorkomen van apparaatbesmetting en voor de nauwkeurigheid van de
temperatuurmetingen van de patiënt.
De reden dat de GENIUS 2 infrarode tympanische thermometer en ACCUSYSTEM√-sondekapjes voor de GENIUS 2-thermometer
voldoen aan de verwachtingen van onze klanten, is dat deze twee onderdelen tijdens het fabricageproces zijn geïntegreerd.
De parameters van het fabricageproces van ACCUSYSTEM√-sondekapjes voor de GENIUS 2-thermometer worden streng
gecontroleerd door Tyco Healthcare/Kendall om geen afwijkingen te krijgen en consistente en betrouwbare producten te leveren.
De verschillende thermoplastische materialen staan ook onder streng toezicht en worden met zorg gekozen teneinde GENIUS
2-thermometers te produceren die consistente temperatuurmetingen verrichten. De fabriekskalibratie en uiteindelijke beoordeling
van een acceptabele GENIUS 2-thermometer is afhankelijk van het gebruik van acceptabele ACCUSYSTEM√-sondekapjes voor de
GENIUS 2-thermometer van Tyco Healthcare/Kendall. Het gebruik van generieke sondekapjes of andere sondekapjes die niet door
Status: Approved Doc#: AW10001726 Effective Date: nulldate
10/23/2008 Artwork#s: WN 182563-01 Job#: 16245
4
GENIUS
2 Controle/kalibrator
Gebruiks- en onderhoudshandleiding
Tyco Healthcare/Kendall zijn geproduceerd, wordt niet ondersteund of goedgekeurd door Tyco Healthcare/Kendall. Het gebruik
van niet-goedgekeurde sondekapjes voor de GENIUS 2-thermometer kan de nauwkeurigheid van de GENIUS 2-thermometer
beïnvloeden. Controleer of het ACCUSYSTEM√-logo op de verpakking van de sondekapjes wordt weergegeven zodat u erop kunt
vertrouwen dat uw thermometer een nauwkeurige temperatuurmeting geeft bij gebruik
Deel IV - Vereist materiaal
1. GENIUS 2-controle/kalibrator
2. Stroommodule met stroomadapter
3. Sondekabel
4. Gebruikshandleiding
5. Te testen GENIUS 2-thermometers
6. GENIUS 2-sondekapjes voor eenmalig gebruik
7. Kleine kruiskopschroevendraaier
Deel V - Systeeminstallatie
De volgende handelingen moeten worden uitgevoerd voor gebruik van de GENIUS 2-controle/kalibrator:
1. Sluit de juiste stroomadapter voor uw land aan op de stroommodule.
2. Sluit de kabel van de stroommodule aan op de stroomaansluiting op het voorpaneel.
3. Sluit de stroommodule aan op een stopcontact.
4. Schakel de eenheid in met de schakelaar op het voorpaneel.
5. Laat de GENIUS 2-controle/kalibrator voor gebruik ten minste 15 minuten opwarmen.
6. Leg een cassette klaar met nieuwe GENIUS 2-sondekapjes.
7. Laat de GENIUS 2-thermometers op kamertemperatuur komen.
8. Prepareer de GENIUS 2-thermometers door de lenzen voor gebruik grondig te reinigen. Raadpleeg de gebruikshandleiding van
de Genius 2-thermometer voor reinigingsinstructies.
9. Verwijder het deurtje van het batterijgedeelte met een kleine kruiskopschroevendraaier. Verwijder alle drie AAA-batterijen.
10. Sluit de ronde connector van de sondekabel aan op het voorpaneel van de GENIUS 2-controle/kalibrator.
11. Sluit het andere uiteinde van de sondeconnector aan op het batterijgedeelte. De connector klikt op zijn plaats op de plek van
de batterijen. Zie afbeelding 2
Status: Approved Doc#: AW10001726 Effective Date: nulldate
10/23/2008 Artwork#s: WN 182563-01 Job#: 16245
5
GENIUS
2 Controle/kalibrator
Gebruiks- en onderhoudshandleiding
Afbeelding 2 - Plaatsen van de sondeconnector in het GENIUS 2 batterijgedeelte
Deel VI - Gebruiksprocedure
De volgende procedure beschrijft de vereiste stappen voor het werken met de GENIUS 2-controle/kalibrator. Om beter bekend te
raken met het product, raden wij aan om dit deel van de handleiding te gebruiken tijdens de kalibratiecontroles van de eerste paar
thermometers.
Installatie van USB-flashstation (optioneel)
Installeer een USB-ashstation in de USB-poort op het voorpaneel indien er een rapportbestand naar het USB-ashstation moet
worden geschreven. Indien het USB-ashstation is geïnstalleerd na de kalibratiecontrole, kan een foutmelding worden weergegeven
bij het schrijven van een rapportbestand als de operator niet wacht totdat het USB-ashstation klaar is met initialiseren. Als de
GENIUS 2-controle/kalibrator is uitgeschakeld, zal al het interne geheugen (dat nog niet is opgeslagen op het USB-ashstation)
verloren gaan.
Stroom inschakelen
Controleer of de stroomadapter van de stroommodule is aangesloten op een geaard stopcontact. Sluit de stroommodule aan op de
stroomconnector op het voorpaneel. Schakel de stroom in naar de GENIUS 2-controle/kalibrator met de stroomschakelaar op het
voorpaneel.
Opstartscherm
Het eerste scherm op de LCD-display is het initialisatiescherm waarin de softwarerevisie en de datum van de GENIUS 2-controle/
kalibrator worden weergegeven. Als zich een intern probleem voordoet tijdens het initialiseren, wordt een foutmelding
weergegeven (zie deel VII).
Herinneringscherm
Het volgende scherm op de LCD-display is een herinnering dat de GENIUS 2-controle/kalibrator voor gebruik ten minste 15 minuten
moet opwarmen na het inschakelen. Op deze manier kunnen de infrarode blackbody-doelen opwarmen tot de juiste temperaturen.
Druk op de codeerknop om dit scherm te sluiten.
Hoofdmenu
Vervolgens wordt het hoofdmenu weergegeven. Dit scherm bevat vier menuopties:
Status: Approved Doc#: AW10001726 Effective Date: nulldate
10/23/2008 Artwork#s: WN 182563-01 Job#: 16245
6
GENIUS
2 Controle/kalibrator
Gebruiks- en onderhoudshandleiding
Kalibratie controleren
Voorkeuren
Gebruiker/locatie-informatie instellen
Tijd/datum instellen
Gebruik de codeerknop om naar de gewenste menuoptie te gaan en deze te markeren. Druk vervolgens op de codeerknop.
Voorkeuren
In het menu Voorkeuren kunt u de taalkeuze instellen. Druk op de codeerknop om het scherm Taalselectie te openen.
Taalkeuze
In het scherm Taalselectie worden de beschikbare talen voor de LCD-display weergegeven. Markeer de gewenste taal om deze te
selecteren. Druk op de codeerknop om de selectie te voltooien en terug te keren naar het hoofdmenu. De geselecteerde taal wordt
opgeslagen in een niet-vluchtig geheugen en wordt gebruikt wanneer het apparaat opnieuw wordt ingeschakeld.
Gebruiker/locatie-informatie instellen
In het scherm Gebruiker/locatie-informatie kunnen de naam van de gebruiker en de gegevens van de organisatie en/of locatie
worden ingevoerd. Deze informatie wordt opgenomen in het testrapportbestand.
Gebruik de codeerknop om door het alfabet te gaan en een bepaalde letterpositie in de gebruikersnaam of locatie-informatie
te kiezen. Als de juiste letter wordt weergegeven, drukt op de codeerknop om de letter te selecteren en verder te gaan naar de
volgende letterpositie. Om een spatie in te voeren, drukt u op de codeerknop wanneer een spatie of lege letter wordt weergegeven
op die letterpositie. De gebruikersnaam kan uit maximaal 20 letters bestaan. Gebruik de codeerknop om het veld voor de organisatie
te selecteren en voer op dezelfde manier een naam voor de locatie in. Ga naar de tekst GEREED en druk op de codeerknop om het
menu te sluiten.
Tijd en datum instellen
De menuoptie Tijd/datum instellen wordt gebruikt om de datum en de tijd in te stellen die worden gebruikt in het testrapport. De
tijd en datum worden intern opgeslagen en automatisch bijgewerkt als de eenheid wordt ingeschakeld.
Gebruik de codeerknop om de uren en minuten te selecteren. Let op: de klok heeft een 24-uurs opmaak waarbij 1 uur ‘s middags
wordt ingevoerd als 13:00, enzovoort. Als de waarde juist is, drukt u op de codeerknop om naar het volgende tijdveld en vervolgens
naar het datumveld te gaan. Voor het datumveld gebruikt u de notatie MMDDJJJJ. Wanneer het laatste datumveld is ingevoerd, sluit
u het menu en keert u terug naar het hoofdmenu door op de codeerknop te drukken. Let op: de tijd en datum worden onder aan het
hoofdmenu weergegeven.
Kalibratie controleren
Met deze menuoptie wordt de kalibratiecontrole van een GENIUS 2-thermometer uitgevoerd en wordt de thermometer, indien
gewenst, automatisch opnieuw gekalibreerd door middel van enkele stapsgewijze schermen. Controleer, voordat u deze menuoptie
gebruikt, of het benodigde materiaal aanwezig is dat in Deel IV wordt genoemd. Hieronder valt ook een cassette met ongebruikte
(nieuwe) sondekapjes.
Stap 1:
Infrarode blackbody-doelen opwarmen. Dit scherm wordt weergegeven terwijl de temperatuur van de doelen wordt gecontroleerd.
Zowel het lage als het hoge doel worden gecontroleerd. Wanneer deze de juiste temperatuur hebben, wordt “OK” weergegeven op
het scherm voor elk doel. U wordt gevraagd op de codeerknop te drukken om verder te gaan naar het volgende scherm. Er verschijnt
een foutmelding als een van de doelen de juiste temperatuur niet heeft bereikt. Controleer in dit geval of de GENIUS 2-controle/
kalibrator zich binnen het juiste bereik van de omgevingstemperatuur bevindt en ten minste 15 minuten heeft kunnen opwarmen
nadat de stroom is ingeschakeld.
Status: Approved Doc#: AW10001726 Effective Date: nulldate
10/23/2008 Artwork#s: WN 182563-01 Job#: 16245
7
GENIUS
2 Controle/kalibrator
Gebruiks- en onderhoudshandleiding
Stap 2:
De GENIUS 2-thermometer inspecteren die wordt getest. Controleer of de thermometerlens niet is beschadigd en of deze geen
oorsmeer en/of ander vuil bevat. Als de thermometerlens vuil is, reinigt u de lens volgens de instructies in de gebruikshandleiding
voor de GENIUS 2-thermometer. Wanneer de thermometerlens is gereinigd, sluit u de GENIUS 2-thermometer aan op de
kalibrator door het ronde uiteinde van de interfacekabel van de thermometer aan te sluiten op de ronde connectoraansluiting
op het voorpaneel van de GENIUS 2-controle/kalibrator (zie afbeelding 3). Open het batterijklepje van de thermometer met een
kleine kruiskopschroevendraaier. Verwijder de batterijen en leg deze apart. Sluit het andere uiteinde van de sondekabel van de
thermometer aan op het batterijgedeelte. Zorg ervoor dat de connector stevig op zijn plaats “klikt”. Zie afbeelding 2 in Deel V.
Let op: telkens wanneer een doel wordt geplaatst, moet een nieuw sondekapje worden gebruikt. Huidvet op eerder gebruikte
sondekapjes of eerder gebruikte sondekapjes die zijn verwarmd door de GENIUS 2-controle/kalibrator kunnen foutieve
metingen veroorzaken waardoor een kalibratiecontrole of herkalibratie kan mislukken. Daarom is het zeer belangrijk dat
elke keer een nieuw sondekapje wordt gebruikt.
Stap 3:
Druk op de codeerknop om verder te gaan en een nieuwe GENIUS 2-sondekapje te installeren op de GENIUS 2-thermometer. Gebruik
een kapje dat in de cassette zit. Richt de sondepunt niet op objecten, zoals handen, vingers of de LCD-display, wanneer het kapje
op de thermometer is geplaatst. Dit kan een onnauwkeurige temperatuuraezing opleveren. Controleer of het sondekapje volledig
is geplaatst (er mag geen ruimte zijn tussen het kapje en de basis van de punt) en de plastic lm geen gaten, scheuren of kreukels
bevat. Als het kapje is geplaatst, plaatst u de sondepunt in het lage temperatuurdoel, naast de knipperende LED. Controleer of de
sondepunt volledig in doelopening is geplaatst. Zie afbeelding 3 voor de juiste richting. Om een time-out te voorkomen, moet u de
sondepunt met kapje in het doel plaatsen binnen 60 seconden vanaf het moment dat u op de codeerknop drukt
GENIUS
F/C˚
F/C˚
Checker / Calibrator
GENIUS
º
F
C
º
Afbeelding 3 - De juiste plaatsing van de GENIUS 2 sondepunt in het doel
Status: Approved Doc#: AW10001726 Effective Date: nulldate
10/23/2008 Artwork#s: WN 182563-01 Job#: 16245
8
GENIUS
2 Controle/kalibrator
Gebruiks- en onderhoudshandleiding
Stap 4:
De plaatsing van de sondepunt in het doel wordt automatisch opgemerkt en op het scherm wordt “Druk op scanknop op GENIUS
2-thermometer” weergegeven. Druk op de scanknop van de GENIUS 2-thermometer terwijl deze in het lage temperatuurdoel is
geplaatst. De GENIUS 2 geeft drie korte geluidssignalen, terwijl de temperatuur van het blackbody-doel wordt gemeten. Als de
temperatuur is opgenomen wordt het volgende scherm weergegeven.
Stap 5:
Verwijder de thermometer uit het lage temperatuurdoel en gooi het sondekapje weg. Op het scherm wordt “Installeer een nieuw
sondekapje en steek de sonde in het hoge doel” weergegeven. Gebruik opnieuw een nieuw GENIUS 2-sondekapje uit de cassette en
zorg ervoor dat u de sondepunt niet richt op objecten, zoals handen, vingers of de LCD-display. Controleer of de sondekapjes volledig
zijn geplaatst (er mag geen ruimte zijn tussen het kapje en de basis van de punt) en de plastic lm geen gaten, scheuren of kreukels
bevat. Als het kapje is geplaatst, plaatst u de sondepunt in het hoge temperatuurdoel, naast de knipperende LED. Controleer of de
sondepunt volledig in doelopening is geplaatst.
Stap 6:
De plaatsing van de sondepunt in het hoge doel wordt automatisch opgemerkt en op het scherm wordt “Druk op scanknop
op GENIUS 2-thermometer” weergegeven. Druk op de scanknop van de GENIUS 2-thermometer terwijl deze in het hoge
temperatuurdoel is geplaatst. De GENIUS 2-thermometer geeft drie korte geluidssignalen, terwijl de temperatuur van het
blackbody-doel wordt gemeten. Als de temperatuur is opgenomen wordt het volgende scherm weergegeven.
Stap 7:
Als de GENIUS 2-thermometer zich binnen bepaalde grenzen bevindt, wordt een resultatenscherm weergegeven met het serienummer
van de thermometer en de bevestiging dat de thermometer is geslaagd voor de kalibratiecontrole. Als de GENIUS 2-thermometer
zich buiten de opgegeven grenzen bevindt, wordt op de display aangegeven dat een kalibratieprocedure wordt uitgevoerd door de
GENIUS 2-controle/kalibrator en dat de gebruiker stappen 3 tot en met 6 nog drie keer moet uitvoeren om herkalibratiegegevens te
kunnen verzamelen. Nadat u stappen 3 tot en met 6 de laatste keer hebt uitgevoerd, voert de GENIUS 2-controle/kalibrator opnieuw
een kalibratie uit van de GENIUS 2-thermometer. Als de herkalibratie is geslaagd, wordt een resultatenscherm weergegeven met
het serienummer van de thermometer en de bevestiging dat de thermometer is geslaagd voor de kalibratie. Als de herkalibratie
niet is geslaagd of als de thermometer niet opnieuw kan worden gekalibreerd, wordt een resultatenscherm weergegeven met het
serienummer van de thermometer en het bericht dat de kalibratie van de thermometer is mislukt.
Stap 8:
Nadat het resultatenscherm is weergegeven, drukt u op de codeerknop om een tweede resultatenscherm weer te geven met de
doeltemperaturen, de metingen van de GENIUS 2 en de afwijkingen tussen het doel en de thermometer.
Stap 9:
In het volgende scherm wordt “Rapport naar USB-ashstation schrijven?” weergegeven, met daaronder Ja en Nee. Als u een rapport
wenst, plaatst u een Flash-station in de USB-poort op het voorpaneel. Let op: het duurt na het plaatsen ongeveer 30 seconden
voordat het USB-ashstation is geïnitialiseerd. Gebruik de codeerknop om naar Ja of Nee te gaan en druk op de codeerknop om een
keuze te maken. Als u Nee selecteert, keert u terug naar het hoofdmenu. Als u Ja selecteert, wordt een rapport geschreven naar het
USB-ashstation, waarbij het serienummer van de thermometer wordt gebruikt als bestandsnaam en .TXT wordt gebruikt als
bestandsextensie. Het bestand bevat de testresultaten en andere informatie, zoals serienummer, datum, tijd, gebruikersnaam en
organisatie. Dit bestand is een eenvoudig ASCII-tekstbestand en is alleen beschikbaar in het Engels. Dit bestand kan worden
gelezen door MicrosoftTM Windows-programma’s zoals Kladblok en kan naar een printer worden verzonden. Nadat het bestand is
geschreven, wordt op het scherm “Druk op knop om verder te gaan” weergegeven. Druk op de codeerknop om terug te keren naar
het hoofdmenu.
Stap 10:
Nadat een kalibratiecontrole/herkalibratie is voltooid, verwijderd u de sondeconnector uit het batterijgedeelte door de connector
heen en weer te bewegen. Plaats vervolgens de drie AAA-batterijen terug en let hierbij op de juiste polariteit. Plaats ten slotte
het deksel van het batterijgedeelte terug en schroef deze vast.
Status: Approved Doc#: AW10001726 Effective Date: nulldate
10/23/2008 Artwork#s: WN 182563-01 Job#: 16245
9
GENIUS
2 Controle/kalibrator
Gebruiks- en onderhoudshandleiding
Stap 11:
Verwijder het USB-ashstation voordat u het deksel van de behuizing sluit. Het deksel dient gesloten te zijn wanneer de GENIUS
2-controle/kalibrator niet in gebruik is
Deel VII - Foutmeldingen
Foutmeldingen zijn voorzien van een nummer, zoals fout 1, enzovoort. Hieronder volgt een uitleg van de fouten:
Fout 1: Interne fout
Deze fout geeft aan dat een probleem is opgetreden in de GENIUS 2-controle/kalibrator. Als deze fout zich blijft voordoen nadat
u het apparaat hebt uitgeschakeld en opnieuw ingeschakeld, dient u de eenheid te retourneren voor onderhoud in de fabriek.
Fout 2: Kan geen verbinding maken met GENIUS 2-thermometer.
Deze fout geeft aan dat er geen communicatie of miscommunicatie plaatsvindt tussen de GENIUS 2-thermometer en de GENIUS
2-controle/kalibrator. Controleer beide uiteinden van de sondekabel van de thermometer, met name het uiteinde dat in het
batterijgedeelte is aangesloten. De pinnen in de sondekabel van de thermometer verliezen mogelijk hun verende eigenschap
wanneer ze zijn gebogen of beschadigd. Controleer de sondekabel en voer de kalibratiecontrole opnieuw uit. Probeer,
indien mogelijk, een andere GENIUS 2-thermometer als het probleem zich blijft voordoen. Indien het probleem niet wordt
opgelost, dient u de eenheid te retourneren voor onderhoud in de fabriek, samen met de thermometerkabel en de GENIUS
2-thermometer waarbij deze fout is opgetreden.
Fout 3: Time-out
Deze fout geeft aan dat de gebruiker te lang heeft gewacht met het uitvoeren van een handeling. Deze fout kan worden
weergegeven bij het aansluiten van de GENIUS 2-thermometer of tijdens het installeren van een sondekapje op de
thermometer. De time-out is ingesteld op één minuut. Voer de handeling waarbij deze fout is opgetreden opnieuw uit.
Fout 4: Doel buiten bereik
Deze fout geeft aan dat een van de blackbody-doelen niet binnen het bereik ligt van de temperatuur. Controleer of de eenheid
wordt gebruikt binnen het bereik van de toegestane omgevingstemperatuur. Raadpleeg Deel VIII. Als de fout zich blijft
voordoen wanneer u het apparaat gebruikt binnen de opgegeven omgevingstemperatuur, zijn de blackbody-doelen mogelijk
beschadigd. Retourneer de eenheid voor onderhoud in de fabriek.
Fout 5: Schrijffout van USB-flashstation
Deze fout geeft aan dat een probleem is opgetreden bij het schrijven van het rapportbestand naar een USB-ashstation.
Controleer of de schakelaar voor schrijfbeveiliging op het USB-station is uitgeschakeld. Controleer of er voldoende vrije ruimte
op het station aanwezig is en of er geen beveiliging of codeerfuncties zijn ingeschakeld.
Fout 6: Omgevingstemperatuur buiten bereik
Deze fout geeft aan dat de omgevingstemperatuur niet binnen het opgegeven bereik ligt voor een nauwkeurige herkalibratie.
Controleer of de eenheid wordt gebruikt binnen het bereik van de toegestane omgevingstemperatuur. Raadpleeg Deel VIII.
Fout 7: Onjuist doel
Deze fout geeft aan dat de GENIUS 2-thermometer in een onjuist doel is geplaatst of te vroeg in het doel is geplaatst. Volg de
instructies op het scherm en voer geen stappen uit voordat de instructies worden weergegeven.
De foutmelding wordt gedurende 60 seconden weergegeven voordat het hoofdmenu opnieuw wordt weergegeven. Druk op de
codeerknop om de foutmelding eerder te sluiten.
.
Status: Approved Doc#: AW10001726 Effective Date: nulldate
10/23/2008 Artwork#s: WN 182563-01 Job#: 16245
10
GENIUS
2 Controle/kalibrator
Gebruiks- en onderhoudshandleiding
Deel VIII - Specificaties van controle/kalibrator
Instelpunten van temperatuur
Laag doel
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .32,2 °C +/- 0,3 °C (90,0 °F +/- 0,5 °F)
Hoog doel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40,6 °C +/- 0,3 °C (105,0 °F +/- 0,5 °F)
Nauwkeurigheid doeltemperatuur. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Initiële grens +/- 0,03 °C (0,05 °F)
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Na 1 jaar +/- 0,06 °C (0,10 °F)
Nauwkeurigheid GENIUS 2-thermometer na herkalibratie*
Metingen tussen 36,0 °C en 39,0 °C (96,8 °F en 102,2 °F) . . . . . . . . . . . . . . . . . . . . . . . . . +/- 0,2 °C (+/- 0,4 °F)
Metingen lager dan 36,0 °C (96,8 °F) of hoger dan 39,0 °C (102,2 °F) . . . . . . . . . . . . . . . . . +/- 0,3 °C (+/- 0,5 °F)
Opwarmtijd voor gebruik . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15 minuten (vanuit een niet-extreme omgeving)
Omgevingstemperatuur in bedrijf . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21,1 - 26,7 °C (70 - 80 °F)
Relatieve vochtigheid in bedrijf . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50 +/- 20% zonder condensvorming
Hoogte. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Tot 2000 meter (6563 feet)
Opslagtemperatuur. . . . . . . . . . . . . . . . . . . . . . . . . . . -25 °C - 55 °C (-13 °F - 131 °F) 85% RL zonder condensvorming
Gemiddelde afmetingen
Lengte
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31,1 cm
Diepte . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15,9 cm
Breedte . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26,0 cm
Gewicht . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3,3 kg
Stroomvoorziening . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100-240 V 0,8-0,4 A 47-63 Hz
Zekeringsvereisten . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Schurter FSF 5 x 20 mm snel smeltend
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Laag uitschakelvermogen 250 V, 3,15 A
Aanbevolen kalibratie-interval van controle/kalibrator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Eén jaar
ESD-immuniteit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Voldoet aan IEC 61000-4-2 Niet-continue Criteria C
Mate van verontreiniging . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Installatiecategorie. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . II
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Alleen voor gebruik binnenshuis
*Nauwkeurigheid na herkalibratie met de GENIUS 2-controle/kalibrator is mogelijk niet gelijk aan fabriekskalibratie
Deel IX - Reinigingsinstructies
LET OP: DOMPEL DE CONTROLE/KALIBRATOR, DE KABELS OF HET NETSNOER NIET ONDER IN WATER OF REINIGINGSOPLOSSINGEN,
REINIG DEZE MET EEN SCHONE, VOCHTIGE (DROGE) DOEK. ALS U ZICH NIET AAN DE IN DIT DOCUMENT BESCHREVEN
REINIGINGSPROCEDURES HOUDT, KAN DIT GEVAAR VOOR DE GEBRUIKERS OPLEVEREN.
Zoals geldt voor elk elektrisch apparaat, dient u erop te letten dat er geen vloeistof in de controle/kalibrator terechtkomt ter
voorkoming van elektrische schokken, brand of schade aan elektrische componenten.
Status: Approved Doc#: AW10001726 Effective Date: nulldate
10/23/2008 Artwork#s: WN 182563-01 Job#: 16245
11
GENIUS
2 Controle/kalibrator
Gebruiks- en onderhoudshandleiding
Algemene reinigingsinstructies
Het reinigen van de GENIUS 2-controle/kalibrator kan als volgt worden uitgevoerd:
HAAL EERST DE STEKKER VAN DE GENIUS 2-CONTROLE/KALIBRATOR UIT HET STOPCONTACT. REINIG HET APPARAAT NOOIT ALS DIT
NOG OP HET ELEKTRICITEITSNET IS AANGESLOTEN.
Voor een algemene reiniging van het buitenoppervlak dient een mild reinigingsmiddel te worden gebruikt. Een vochtige doek
met isopropylalcohol kan worden gebruikt om de eenheid te reinigen. GEBRUIK GEEN schuurmiddelen of oplosmiddelen. Vermijd
overmatige bevochtiging rond de doeleenheden en de elektrische connectoren.
Probeer niet het doel onder de rubberen flappen te reinigen.
Reinig de doelnesten of flappen niet met alcohol.
Bij enige twijfel over de geschiktheid van een reinigingsmiddel voor een onderdeel van dit apparaat of het materiaal, neemt u
contact op met de klantenservice van Tyco Healthcare/Kendall (zie Deel XI).
Sectie X - Probleemoplossing
Hieronder volgen enkele veelvoorkomende problemen:
Probleem Mogelijke oorzaak Correcties
GENIUS 2-kalibrator wordt niet
ingeschakeld
GENIUS 2-kalibrator is niet aangesloten op netstroom Controleer de netsnoeren
Zekering doorgeslagen Vervang de zekering
Testrapport wordt niet naar USB-
station geschreven
USB-flashstation is niet geformatteerd Gebruik een pc of laptop om het Flash-station te formatteren
USB-flashstation is tegen schrijven beveiligd Gebruik een pc of laptop om schrijfbeveiliging uit te schakelen
USB-flashstation beschikt niet over voldoende vrije ruimte Gebruik een pc of laptop om overbodige items te wissen of gebruik een ander
USB-flashstation
USB-flashstation is niet compatibel met GENIUS 2-controle/
kalibrator
Gebruik een ander USB-flashstation
GENIUS 2-kalibrator heeft onvoldoende tijd gehad om het USB-
flashstation te detecteren
Probeer opnieuw te schrijven naar het USB-station
Meerdere GENIUS 2-thermometers
worden niet opnieuw gekalibreerd
Sondekapje zit vast in een van de blackbody-doelen Verwijder het sondekapje uit het blackbody-doel met een klein pincet of een
hemostaat. Beschadig de binnenkant van het doel niet.
Sondekapje is vuil Gebruik een nieuw sondeklepje voor elke meting. Gebruik een sondekapje
niet opnieuw. Sondekapjes mogen niet met de hand worden aangeraakt.
Thermometerlens is vuil Reinig de thermometerlens volgens de instructies in de gebruikshandleiding
voor de GENIUS 2.
GENIUS 2-thermometer is niet juist geplaatst Controleer of de GENIUS 2-thermometer volledig en onder de juiste hoek in
het doel is geplaatst.
GENIUS 2-thermometers, -sondekapjes en/of -controle/kalibrator is
niet op kamertemperatuur.
Controleer of de GENIUS 2-thermometers, -sondekapjes en/of -kalibrator
voldoende tijd hebben gehad om op kamertemperatuur te komen voordat u
verder gaat.
Fout 6 doet zich herhaaldelijk voor
bij het testen van een bepaalde
GENIUS 2-thermometer
Mogelijk probleem met stroomverbruik met GENIUS 2-thermometer Herhaal de kalibratiecontrole/herkalibratie met een andere GENIUS
2-thermometer en retourneer de thermometer waarbij het probleem zich
voordoet voor onderhoud in de fabriek.
GENIUS 2-thermometer lijkt te
hoog of te laag te meten na een
geslaagde herkalibratie
Sondekapje is opnieuw gebruikt na kalibratiecontrole/herkalibratie Herhaal kalibratiecontrole/herkalibratie met een nieuw sondekapje voor elke
plaatsing in het doel tijdens de procedure.
Status: Approved Doc#: AW10001726 Effective Date: nulldate
10/23/2008 Artwork#s: WN 182563-01 Job#: 16245
12
GENIUS
2 Controle/kalibrator
Gebruiks- en onderhoudshandleiding
Sectie XI - Klantenservice
De GENIUS 2-controle/kalibrator moet eenmaal per jaar worden gekalibreerd en gecontroleerd in de fabriek. Neem contact op met
Tyco Healthcare/Kendall en retourneer het apparaat op of voor de datum van kalibratie. Op de controle/kalibrator wordt één maand
voor de datum van kalibratie een herinnering weergegeven.
LET OP: HET APPARAAT BEVAT GEEN ONDERDELEN DIE ONDERHOUDEN KUNNEN WORDEN, VOOR ONDERHOUD RETOURNEREN NAAR DE
FABRIEK.
Indien het nodig is een eenheid voor reparatie te retourneren, gaat u als volgt te werk:
1. Neem contact op met Tyco Healthcare/Kendall zoals hieronder staat aangegeven of met uw vertegenwoordiger van de
klantenservice voor het verkrijgen van een retourautorisatienummer en aanwijzingen voor verzending.
2. Verzend een verzekerd pakket naar uw regionale contactpersoon of de juiste onderstaande locatie.
Verenigde Staten Canada Buiten de VS en Canada
Tyco Healthcare Group LP Tyco Healthcare Canada Tyco Healthcare Group LP
98.6 Faichney Drive 7300 Trans Canada Highway Unit 2 Talisman Business Center, London Road
Watertown, NY 13601, VS Pointe-Claire, QC Bicester, OX26 6HR, VK
+1 800 448-0190 of H9R 1C7 +44-1869-328065
(315) 788-5246 +1 877 664-8926 or
(514) 695-1220
Onderdelenlijst
Neem contact op met uw plaatselijke klantenservice of uw verkoopvertegenwoordiger voor de onderstaande onderdelen.
Beschrijving
GENIUS 2-controle/kalibrator. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .303097
Vernieuwde GENIUS 2-controle/kalibrator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .303096
Sectie XII - Garantie
Tyco Healthcare/Kendall garandeert dat de GENIUS 2-controle/kalibrator vrij zal zijn van gebreken in materiaal en vakmanschap
onder normaal gebruik gedurende een periode van één (1) jaar vanaf datum van levering door Tyco Healthcare/Kendall aan de eerste
koper. De aansprakelijkheid is hierbij beperkt tot vervanging of reparatie, dit ter beoordeling van Tyco Healthcare/Kendall, door
het servicecentrum van TycoHealthcare/Kendall van enig product dat onder normaal gebruik en service volgens Tyco Healthcare/
Kendall gebreken vertoont in materiaal of vakmanschap. Deze garantie is niet van toepassing op en Tyco Healthcare/Kendall zal
niet verantwoordelijk zijn voor enig verlies geleden in verband met de aankoop van enig product dat is gerepareerd door enige
andere partij dan Tyco Healthcare/Kendall of haar geautoriseerde vertegenwoordiger, of dat is blootgesteld aan onjuist gebruik,
verwaarlozing of ongeluk, of waarvan het serienummer is gewijzigd, beschadigd of verwijderd, of dat is gebruikt op een andere
wijze dan volgens de instructies verstrekt door Tyco Healthcare/Kendall. Tyco Healthcare/Kendall accepteert noch volmachtigt enige
vertegenwoordiger of andere partij uit haar naam enige aansprakelijkheid te aanvaarden anders dan de aansprakelijkheid die hier
uitdrukkelijk is vermeld.
ALLE ANDERE AANSPRAKEN OP GARANTIE, EXPLICIET OF IMPLICIET, MET INBEGRIP VAN IMPLICIETE GARANTIES VAN
VERKOOPBAARHEID OF GESCHIKTHEID VOOR EEN BEPAALD DOEL OF EEN BEPAALDE TOEPASSING ANDERS DAN HETGEEN
UITDRUKKELIJK IS GESTELD IN DE BIJ HET PRODUCT BEHORENDE DOCUMENTATIE, WORDEN DOOR Tyco Healthcare/Kendall
AFGEWEZEN. IN GEEN GEVAL ZAL Tyco Healthcare/Kendall AANSPRAKELIJK ZIJN VOOR ENIGE INCIDENTELE, INDIRECTE OF
VERVOLGSCHADE IN VERBAND MET DE AANKOOP OF HET GEBRUIK VAN ZIJN PRODUCTEN
Status: Approved Doc#: AW10001726 Effective Date: nulldate
10/23/2008 Artwork#s: WN 182563-01 Job#: 16245
13
GENIUS
2 Controle/kalibrator
Gebruiks- en onderhoudshandleiding
Sectie XIII - Verklaring elektromagnetische conformiteit
De Genius 2-controle/kalibrator is vervaardigd en getest volgens de normen IEC/EN61010-1 en EN61326-1.
De controle/kalibrator is bedoeld voor gebruik in de onderstaande gespeciceerde elektromagnetische omgeving. De gebruiker van
de kalibrator dient ervoor te zorgen dat het apparaat in een dergelijke omgeving wordt gebruikt
Emissietesten Conformiteit Richtlijn - elektromagnetische omgeving
Geleid en uitgestraald
Emissies (EN61326-1)
Groep 1, Klasse B De kalibratoremissies zijn zeer laag en zullen naar verwachting
geen nabije elektronische apparatuur storen.
Harmonische stroom
(EN61000-3-2)
Klasse A
Spanningsschommelingen en
ikkering (EN61000-3-3)
Wordt aan voldaan
Immuniteit voor uitgestraalde
storing (EN61000-4-3)
Wordt aan voldaan
Immuniteit voor geleide storing
(EN61000-4-6)
Wordt aan voldaan
Immuniteit voor
stroomfrequentie magnetisch
veld (EN61000-4-8)
Wordt aan voldaan
Immuniteit voor
spanningsdalingen en
onderbrekingen (EN61000-4-11)
Wordt aan voldaan
Immuniteit voor elektrische
snelle transiënt/bursts
(EN61000-4-4)
Wordt aan voldaan
Immuniteit plotselinge dalingen
(EN61000-4-5)
Wordt aan voldaan
Elektrostatische ontlading
(EN61000-4-2))
Voldoet aan criteria C,
Niet-continu
Status: Approved Doc#: AW10001726 Effective Date: nulldate
10/23/2008 Artwork#s: WN 182563-01 Job#: 16245

Documenttranscriptie

Status: Approved Doc#: AW10001726 Effective Date: nulldate Inhoudsopgave Nederlands Pagina Deel I - Functiebeschrijving ............................................................................................................................................................. 1 Deel II — Veiligheid en waarschuwingen..................................................................................................................................... 2 Sectie III - Pictogrammen en hun betekenis ................................................................................................................................. 3 Deel IV - Vereist materiaal ............................................................................................................................................................... 4 Deel V — Systeeminstallatie .......................................................................................................................................................... 4 Deel VI - Gebruiksprocedure ............................................................................................................................................................ 5 Stroom inschakelen ....................................................................................................................................................................... 5 Opstartscherm............................................................................................................................................................................... 5 Herinneringscherm ....................................................................................................................................................................... 5 Hoofdmenu ................................................................................................................................................................................... 5 Voorkeuren .................................................................................................................................................................................... 5 Taalkeuze ...................................................................................................................................................................................... 5 Gebruiker/locatie-informatie instellen ......................................................................................................................................... 6 Tijd en datum instellen .................................................................................................................................................................... 6 Kalibratie controleren ...................................................................................................................................................................... 6 Deel VII - Foutmeldingen ................................................................................................................................................................. 9 Deel VIII - Specificaties van controle/kalibrator ......................................................................................................................... 10 Deel IX - Reinigingsinstructies ...................................................................................................................................................... 10 Sectie X - Probleemoplossing ........................................................................................................................................................ 11 Sectie XI - Klantenservice ............................................................................................................................................................... 12 Sectie XII — Garantie ..................................................................................................................................................................... 12 Sectie XIII — Verklaring elektromagnetische conformiteit .................................................................................................... 13 Dit product bevat software waarvan Tyco Healthcare Group LP de enige eigenaar is. Tyco Healthcare Group LP verleent de gebruiker een niet-exclusieve en beperkte licentie om de software te gebruiken in overeenstemming met de bedieningsinstructies. Een kopie van de licentie kan worden verkregen bij Tyco Healthcare Group LP. 10/23/2008 Artwork#s: WN 182563-01 Job#: 16245 Status: Approved Doc#: AW10001726 Effective Date: nulldate Deel I - Functiebeschrijving Deze handleiding beschrijft de werking van de GENIUS 2-controle/kalibrator. Dit apparaat is ontwikkeld om de kalibratie van de GENIUS 2 tympanische thermometer te controleren en indien nodig automatisch de thermometer opnieuw te kalibreren. Daarnaast beschikt de GENIUS 2-controle/kalibrator over de mogelijkheid om testrapporten te genereren voor geteste thermometers en deze op een USB-flashstation op te slaan. De GENIUS 2-controle/kalibrator kan gebruikt worden voor alle softwarerevisies van de GENIUS 2-thermometer. De GENIUS 2-controle/kalibrator bevat twee onafhankelijk gecontroleerde infrarode kalibratiedoelen die lijken op de kalibratiedoelen in de fabriek. Deze “blackbody”-doelen zijn speciaal ontworpen voor een efficiënte overdracht van stralingswarmte en geven infrarode straling af die het menselijke oor benadert bij temperaturen van 32,2 °C (90,0 °F) en 40,6 °C (105,0 °F). 1 2 3 4 5 F 3.15A 250V 11 F/C˚ GENIUS™ Checker / Calibrator 6 F/C˚ ™ 10 9 7 8 Afbeelding 1 - Layout van voorpaneel Legende 1 2 3 4 5 6 Voedingaansluiting Stroomschakelaar Zekeringhouder LED van lage temperatuurdoel Lage temperatuurdoel Connector van GENIUS 2 thermometer GENIUS™ 2 Controle/kalibrator 7 8 9 10 11 Hoge temperatuurdoel LED van hoge temperatuurdoel LCD-display Codeerknop USB-poort 1 10/23/2008 Artwork#s: WN 182563-01 Job#: 16245 Gebruiks- en onderhoudshandleiding Status: Approved Doc#: AW10001726 Effective Date: nulldate Deel II - Veiligheid en waarschuwingen • De GENIUS 2-controle/kalibrator is bedoeld voor gebruik zoals wordt aangegeven in deze handleiding. De bescherming door de veiligheidsmaterialen in de GENIUS 2-controle/kalibrator kunnen teniet worden gedaan als het apparaat niet wordt gebruikt volgens de aanwijzingen in deze handleiding. • De behuizing van de GENIUS 2-controle/kalibrator is niet waterdicht. Vermijd het morsen van enige vloeistof op het voorpaneel van het apparaat. De GENIUS 2-controle/kalibrator wordt gevoed via het elektriciteitsnet. Vermijd een mogelijk contact tussen vloeistof en enig buitenoppervlak van het apparaat wanneer deze is aangesloten op het net. • De oppervlaktekwaliteit van de binnenzijde van de referentie blackbody-doelen is zeer belangrijk voor de werking als doeltreffende infrarode radiatoren. Men moet voorzichtig zijn bij het werken in de nabijheid van de blackbody-doelen, om zeker te stellen dat er geen contact wordt gemaakt met het oppervlak binnen de rubberen flappen. • In het geval van een gesprongen zekering dient u de zekering te vervangen met een zekering die precies overeenkomt met de nominale spanning, stroom en vormfactor, zoals aangegeven in het gedeelte Specificaties van GENIUS 2-controle/kalibrator van deze handleiding. Controleer of het apparaat is losgekoppeld van het elektriciteitsnet voordat een gesprongen zekering wordt vervangen. • Het apparaat mag niet worden gebruikt in de aanwezigheid van brandbare anesthetica. Er bestaat explosiegevaar in de nabijheid van deze anesthetica. • Dit apparaat is niet voorzien van onderdelen die door de gebruiker kunnen worden onderhouden. Het voorpaneel niet verwijderen. Laat het onderhoud uitvoeren door gekwalificeerd onderhoudspersoneel. • Gebruik alleen de geleverde stroommodule bij dit apparaat. • Wanneer u wordt gevraagd om de GENIUS 2-thermometer in een van de blackbody-doelen te plaatsen, heeft u 60 seconden de tijd om deze handeling uit te voeren. De procedure zal worden afgebroken als deze periode wordt overschreden. • Zorg ervoor dat de thermometerlens schoon is voordat u een GENIUS 2-thermometer controleert en/of kalibreert. Wanneer deze effectief is gereinigd, bent u ervan verzekerd dat de toegestane kalibratieafwijking correct wordt gecontroleerd. Raadpleeg de reinigingsinstructies in de gebruikshandleiding van de Genius 2-thermometer. • Bij het installeren van een GENIUS 2-sondekapje op de punt van de GENIUS 2-thermometer, dient u alleen kapjes te gebruiken die in een cassette zitten en die niet los liggen. Probeer elk handcontact met de sondekapjes te vermijden, om onnauwkeurige aflezingen te voorkomen. Gebruik de sondekapjes niet opnieuw. Gebruik alleen GENIUS 2-sondekapjes. • Laat de GENIUS 2-kalibrator na het inschakelen gedurende ten minste 15 minuten opwarmen voor gebruik. Controleer of de GENIUS 2-thermometers, sondekapjes en de GENIUS 2-controle/kalibrator voldoende tijd hebben gehad om op kamertemperatuur te komen voordat u verder gaat. Zorg voor een langere opwarmperiode als het apparaat of de thermometers een temperatuur hebben op de uiterste grenzen van de toegestane omgevingstemperatuur. De draagbaarheid van de GENIUS 2-controle/kalibrator staat verplaatsing toe van het apparaat naar ruimten die voldoen aan de vereisten voor de omgevingstemperatuur. • Gebruik een USB-flashstation die is voorzien van het CE-keurmerk en die voldoet aan de elektromagnetische stralingsnorm EN55022. • Plaats de GENIUS 2-thermometer niet op het oppervlak van de GENIUS 2-controle/kalibrator. De temperatuursensoren in de sondetip van de thermometer zullen de warmte van de “blackbody”-doelen waarnemen via het oppervlak van het voorpaneel. Dit zorgt voor een instabiliteit van de thermometersensoren, wat zal leiden tot onjuiste temperatuuraflezingen. • De GENIUS 2-controle/kalibrator mag alleen binnenshuis worden gebruikt, in een ruimte zonder tocht en zonder grote temperatuurswisselingen. • De USB-poort op het voorpaneel is alleen bedoeld voor gebruik met een USB-flashstation. Er mogen geen andere USB-apparaten op de USB-poort worden aangesloten. De USB-flashstation moet voor het inschakelen van het apparaat zijn geïnstalleerd. • Retourneer de GENIUS 2-controle/kalibrator aan de fabriek voor herkalibratie indien het apparaat is gevallen GENIUS™ 2 Controle/kalibrator 2 10/23/2008 Artwork#s: WN 182563-01 Job#: 16245 Gebruiks- en onderhoudshandleiding Status: Approved Doc#: AW10001726 Effective Date: nulldate Sectie III - Pictogrammen en hun betekenis MARK SYMBOLS 0123 SN CE MARK SYMBOLS Single Use TERILE R Serienummer Zie bijgaande documenten 0123 STERILE SingleEO Use TERILE A STERILE STERILE R STERILE EO CE-markering LOT Weggooien in overeenstemming met richtlijnen voor elektrisch en elektronisch afval 40° 0° Extremes FRAGILE 2 STERILIZE NON STERILE Keep Away From Sunlight F/C˚ ospheric pressure limitation Use By Hoge temperatuurdoelAvoid Temperature F/C˚ THIS SIDE UP Dubbele of versterkte isolatiebescherming LOT Instructions For Use Avoid Temperature Extremes STERILE FRAGILE Use By USB-poort0° R STERILE EO TERILE ASTERILESTERILE DO NOT RESTERILIZE LISTED Keep Dry 2 UL-opname-informatie (LABORATORY EQUIPMENT) STERILIZE NON 46EA STERILE Lage temperatuurdoel Keep Away Keep Dry From Sunlight Humidity Limitation DO NOT RESTERILIZE Let op: Alleen voor gebruik binnenshuis PROTECT FROM HEAT AND Aansluiting van GENIUS RADIOACTIVE SOURCES 2-thermometer N-CE MARK SYMBOLS Atmospheric pressure limitation EX– FREE Consult Stand-bystroom Caution, consultdoor Vervangen Consult accompanying Instructions gelijksoortige zekering 40° documents For Use STERILE A STERILE TERILE R STERILE EO THIS SIDE UP Fabricagedatum Do not use if package is opened or damaged. Caution, consult accompanying documents STERILE A Fabrikant Do not use if package is opened or damaged. Humidity Limitation ™ PROTECT FROM HEAT AND RADIOACTIVE SOURCES NON-CE MARK SYMBOLS De GENIUS 2 infrarode tympanische thermometer is een betrouwbaar en nauwkeurig apparaat voor het opnemen van de Corrugated Recycles temperatuur. De GENIUS 2-thermometer dankt zijn nauwkeurigheid aan het ontwerp, de gecontroleerde kalibratieprocedures en de strikte fabricagemethoden. Een van de belangrijkste functionele onderdelen van een thermometer is het ACCUSYSTEM√-sondekapje POWDER geproduceerd door Tyco Healthcare/Kendall. Het ACCUSYSTEM√-sondekapje voor de GENIUS 2-thermometer fungeert, wanneer Corrugated Recycles geplaatst op een GENIUS 2-thermometer, als een infectiepreventiedrempel tussen de patiënt en het apparaat en als medium voor DEHP– FREE LATEX– FREE warmtetransmissie van de patiënt naar de thermometer. De functionaliteit van het ACCUSYSTEM√-sondekapje voor de NON-PYROGENIC POWDER– FREE POWDER MR UNSAFE GENIUS 2-thermometer is zeer belangrijk voor het voorkomen van apparaatbesmetting en voor de nauwkeurigheid van de temperatuurmetingen van de patiënt. DEHP– FREE MR MR ECTRONIC SYMBOLS AND CERTIFIED TESTING MARKS NON-PYROGENIC POWDER– FREE UNSAFE De redenMRdat de GENIUS 2 infrarode tympanische thermometer en ACCUSYSTEM√-sondekapjes voor de GENIUS 2-thermometer voldoen aan de verwachtingen van onze klanten, is dat deze twee onderdelen tijdens het fabricageproces zijn geïntegreerd. ELECTRONIC SYMBOLS AND TESTING MARKS De parameters vanCERTIFIED het fabricageproces van ACCUSYSTEM√-sondekapjes voor de GENIUS 2-thermometer worden streng gecontroleerd door Tyco Healthcare/Kendall om geen afwijkingen te krijgen en consistente en betrouwbare producten te leveren. De verschillende thermoplastische materialen staan ook onder streng toezicht en worden met zorg gekozen teneinde GENIUS 2-thermometers te produceren die consistente temperatuurmetingen verrichten. De fabriekskalibratie en uiteindelijke beoordeling van een acceptabele GENIUS 2-thermometer is afhankelijk van het gebruik van acceptabele ACCUSYSTEM√-sondekapjes voor de GENIUS 2-thermometer van Tyco Healthcare/Kendall. Het gebruik van generieke sondekapjes of andere sondekapjes die niet door US C GENIUS™ 2 Controle/kalibrator US 3 10/23/2008 Artwork#s: WN 182563-01 Job#: 16245 Gebruiks- en onderhoudshandleiding Status: Approved Doc#: AW10001726 Effective Date: nulldate Tyco Healthcare/Kendall zijn geproduceerd, wordt niet ondersteund of goedgekeurd door Tyco Healthcare/Kendall. Het gebruik van niet-goedgekeurde sondekapjes voor de GENIUS 2-thermometer kan de nauwkeurigheid van de GENIUS 2-thermometer beïnvloeden. Controleer of het ACCUSYSTEM√-logo op de verpakking van de sondekapjes wordt weergegeven zodat u erop kunt vertrouwen dat uw thermometer een nauwkeurige temperatuurmeting geeft bij gebruik Deel IV - Vereist materiaal 1. GENIUS 2-controle/kalibrator 2. Stroommodule met stroomadapter 3. Sondekabel 4. Gebruikshandleiding 5. Te testen GENIUS 2-thermometers 6. GENIUS 2-sondekapjes voor eenmalig gebruik 7. Kleine kruiskopschroevendraaier Deel V - Systeeminstallatie De volgende handelingen moeten worden uitgevoerd voor gebruik van de GENIUS 2-controle/kalibrator: 1. 2. 3. 4. 5. 6. 7. 8. Sluit de juiste stroomadapter voor uw land aan op de stroommodule. Sluit de kabel van de stroommodule aan op de stroomaansluiting op het voorpaneel. Sluit de stroommodule aan op een stopcontact. Schakel de eenheid in met de schakelaar op het voorpaneel. Laat de GENIUS 2-controle/kalibrator voor gebruik ten minste 15 minuten opwarmen. Leg een cassette klaar met nieuwe GENIUS 2-sondekapjes. Laat de GENIUS 2-thermometers op kamertemperatuur komen. Prepareer de GENIUS 2-thermometers door de lenzen voor gebruik grondig te reinigen. Raadpleeg de gebruikshandleiding van de Genius 2-thermometer voor reinigingsinstructies. 9. Verwijder het deurtje van het batterijgedeelte met een kleine kruiskopschroevendraaier. Verwijder alle drie AAA-batterijen. 10. Sluit de ronde connector van de sondekabel aan op het voorpaneel van de GENIUS 2-controle/kalibrator. 11. Sluit het andere uiteinde van de sondeconnector aan op het batterijgedeelte. De connector klikt op zijn plaats op de plek van de batterijen. Zie afbeelding 2 GENIUS™ 2 Controle/kalibrator 4 10/23/2008 Artwork#s: WN 182563-01 Job#: 16245 Gebruiks- en onderhoudshandleiding Status: Approved Doc#: AW10001726 Effective Date: nulldate Afbeelding 2 - Plaatsen van de sondeconnector in het GENIUS 2 batterijgedeelte Deel VI - Gebruiksprocedure De volgende procedure beschrijft de vereiste stappen voor het werken met de GENIUS 2-controle/kalibrator. Om beter bekend te raken met het product, raden wij aan om dit deel van de handleiding te gebruiken tijdens de kalibratiecontroles van de eerste paar thermometers. Installatie van USB-flashstation (optioneel) Installeer een USB-flashstation in de USB-poort op het voorpaneel indien er een rapportbestand naar het USB-flashstation moet worden geschreven. Indien het USB-flashstation is geïnstalleerd na de kalibratiecontrole, kan een foutmelding worden weergegeven bij het schrijven van een rapportbestand als de operator niet wacht totdat het USB-flashstation klaar is met initialiseren. Als de GENIUS 2-controle/kalibrator is uitgeschakeld, zal al het interne geheugen (dat nog niet is opgeslagen op het USB-flashstation) verloren gaan. Stroom inschakelen Controleer of de stroomadapter van de stroommodule is aangesloten op een geaard stopcontact. Sluit de stroommodule aan op de stroomconnector op het voorpaneel. Schakel de stroom in naar de GENIUS 2-controle/kalibrator met de stroomschakelaar op het voorpaneel. Opstartscherm Het eerste scherm op de LCD-display is het initialisatiescherm waarin de softwarerevisie en de datum van de GENIUS 2-controle/ kalibrator worden weergegeven. Als zich een intern probleem voordoet tijdens het initialiseren, wordt een foutmelding weergegeven (zie deel VII). Herinneringscherm Het volgende scherm op de LCD-display is een herinnering dat de GENIUS 2-controle/kalibrator voor gebruik ten minste 15 minuten moet opwarmen na het inschakelen. Op deze manier kunnen de infrarode blackbody-doelen opwarmen tot de juiste temperaturen. Druk op de codeerknop om dit scherm te sluiten. Hoofdmenu Vervolgens wordt het hoofdmenu weergegeven. Dit scherm bevat vier menuopties: GENIUS™ 2 Controle/kalibrator 5 10/23/2008 Artwork#s: WN 182563-01 Job#: 16245 Gebruiks- en onderhoudshandleiding Status: Approved Doc#: AW10001726 Effective Date: nulldate Kalibratie controleren Voorkeuren Gebruiker/locatie-informatie instellen Tijd/datum instellen Gebruik de codeerknop om naar de gewenste menuoptie te gaan en deze te markeren. Druk vervolgens op de codeerknop. Voorkeuren In het menu Voorkeuren kunt u de taalkeuze instellen. Druk op de codeerknop om het scherm Taalselectie te openen. Taalkeuze In het scherm Taalselectie worden de beschikbare talen voor de LCD-display weergegeven. Markeer de gewenste taal om deze te selecteren. Druk op de codeerknop om de selectie te voltooien en terug te keren naar het hoofdmenu. De geselecteerde taal wordt opgeslagen in een niet-vluchtig geheugen en wordt gebruikt wanneer het apparaat opnieuw wordt ingeschakeld. Gebruiker/locatie-informatie instellen In het scherm Gebruiker/locatie-informatie kunnen de naam van de gebruiker en de gegevens van de organisatie en/of locatie worden ingevoerd. Deze informatie wordt opgenomen in het testrapportbestand. Gebruik de codeerknop om door het alfabet te gaan en een bepaalde letterpositie in de gebruikersnaam of locatie-informatie te kiezen. Als de juiste letter wordt weergegeven, drukt op de codeerknop om de letter te selecteren en verder te gaan naar de volgende letterpositie. Om een spatie in te voeren, drukt u op de codeerknop wanneer een spatie of lege letter wordt weergegeven op die letterpositie. De gebruikersnaam kan uit maximaal 20 letters bestaan. Gebruik de codeerknop om het veld voor de organisatie te selecteren en voer op dezelfde manier een naam voor de locatie in. Ga naar de tekst GEREED en druk op de codeerknop om het menu te sluiten. Tijd en datum instellen De menuoptie Tijd/datum instellen wordt gebruikt om de datum en de tijd in te stellen die worden gebruikt in het testrapport. De tijd en datum worden intern opgeslagen en automatisch bijgewerkt als de eenheid wordt ingeschakeld. Gebruik de codeerknop om de uren en minuten te selecteren. Let op: de klok heeft een 24-uurs opmaak waarbij 1 uur ‘s middags wordt ingevoerd als 13:00, enzovoort. Als de waarde juist is, drukt u op de codeerknop om naar het volgende tijdveld en vervolgens naar het datumveld te gaan. Voor het datumveld gebruikt u de notatie MMDDJJJJ. Wanneer het laatste datumveld is ingevoerd, sluit u het menu en keert u terug naar het hoofdmenu door op de codeerknop te drukken. Let op: de tijd en datum worden onder aan het hoofdmenu weergegeven. Kalibratie controleren Met deze menuoptie wordt de kalibratiecontrole van een GENIUS 2-thermometer uitgevoerd en wordt de thermometer, indien gewenst, automatisch opnieuw gekalibreerd door middel van enkele stapsgewijze schermen. Controleer, voordat u deze menuoptie gebruikt, of het benodigde materiaal aanwezig is dat in Deel IV wordt genoemd. Hieronder valt ook een cassette met ongebruikte (nieuwe) sondekapjes. Stap 1: Infrarode blackbody-doelen opwarmen. Dit scherm wordt weergegeven terwijl de temperatuur van de doelen wordt gecontroleerd. Zowel het lage als het hoge doel worden gecontroleerd. Wanneer deze de juiste temperatuur hebben, wordt “OK” weergegeven op het scherm voor elk doel. U wordt gevraagd op de codeerknop te drukken om verder te gaan naar het volgende scherm. Er verschijnt een foutmelding als een van de doelen de juiste temperatuur niet heeft bereikt. Controleer in dit geval of de GENIUS 2-controle/ kalibrator zich binnen het juiste bereik van de omgevingstemperatuur bevindt en ten minste 15 minuten heeft kunnen opwarmen nadat de stroom is ingeschakeld. GENIUS™ 2 Controle/kalibrator 6 10/23/2008 Artwork#s: WN 182563-01 Job#: 16245 Gebruiks- en onderhoudshandleiding Status: Approved Doc#: AW10001726 Effective Date: nulldate Stap 2: De GENIUS 2-thermometer inspecteren die wordt getest. Controleer of de thermometerlens niet is beschadigd en of deze geen oorsmeer en/of ander vuil bevat. Als de thermometerlens vuil is, reinigt u de lens volgens de instructies in de gebruikshandleiding voor de GENIUS 2-thermometer. Wanneer de thermometerlens is gereinigd, sluit u de GENIUS 2-thermometer aan op de kalibrator door het ronde uiteinde van de interfacekabel van de thermometer aan te sluiten op de ronde connectoraansluiting op het voorpaneel van de GENIUS 2-controle/kalibrator (zie afbeelding 3). Open het batterijklepje van de thermometer met een kleine kruiskopschroevendraaier. Verwijder de batterijen en leg deze apart. Sluit het andere uiteinde van de sondekabel van de thermometer aan op het batterijgedeelte. Zorg ervoor dat de connector stevig op zijn plaats “klikt”. Zie afbeelding 2 in Deel V. Let op: telkens wanneer een doel wordt geplaatst, moet een nieuw sondekapje worden gebruikt. Huidvet op eerder gebruikte sondekapjes of eerder gebruikte sondekapjes die zijn verwarmd door de GENIUS 2-controle/kalibrator kunnen foutieve metingen veroorzaken waardoor een kalibratiecontrole of herkalibratie kan mislukken. Daarom is het zeer belangrijk dat elke keer een nieuw sondekapje wordt gebruikt. Stap 3: Druk op de codeerknop om verder te gaan en een nieuwe GENIUS 2-sondekapje te installeren op de GENIUS 2-thermometer. Gebruik een kapje dat in de cassette zit. Richt de sondepunt niet op objecten, zoals handen, vingers of de LCD-display, wanneer het kapje op de thermometer is geplaatst. Dit kan een onnauwkeurige temperatuuraflezing opleveren. Controleer of het sondekapje volledig is geplaatst (er mag geen ruimte zijn tussen het kapje en de basis van de punt) en de plastic film geen gaten, scheuren of kreukels bevat. Als het kapje is geplaatst, plaatst u de sondepunt in het lage temperatuurdoel, naast de knipperende LED. Controleer of de sondepunt volledig in doelopening is geplaatst. Zie afbeelding 3 voor de juiste richting. Om een time-out te voorkomen, moet u de sondepunt met kapje in het doel plaatsen binnen 60 seconden vanaf het moment dat u op de codeerknop drukt S™ NIU ºC º F GE F/ C˚ S™ IU ali EN /C Ch ec ke r G br ato r F/ C˚ ™ Afbeelding 3 - De juiste plaatsing van de GENIUS 2 sondepunt in het doel GENIUS™ 2 Controle/kalibrator 7 10/23/2008 Artwork#s: WN 182563-01 Job#: 16245 Gebruiks- en onderhoudshandleiding Status: Approved Doc#: AW10001726 Effective Date: nulldate Stap 4: De plaatsing van de sondepunt in het doel wordt automatisch opgemerkt en op het scherm wordt “Druk op scanknop op GENIUS 2-thermometer” weergegeven. Druk op de scanknop van de GENIUS 2-thermometer terwijl deze in het lage temperatuurdoel is geplaatst. De GENIUS 2 geeft drie korte geluidssignalen, terwijl de temperatuur van het blackbody-doel wordt gemeten. Als de temperatuur is opgenomen wordt het volgende scherm weergegeven. Stap 5: Verwijder de thermometer uit het lage temperatuurdoel en gooi het sondekapje weg. Op het scherm wordt “Installeer een nieuw sondekapje en steek de sonde in het hoge doel” weergegeven. Gebruik opnieuw een nieuw GENIUS 2-sondekapje uit de cassette en zorg ervoor dat u de sondepunt niet richt op objecten, zoals handen, vingers of de LCD-display. Controleer of de sondekapjes volledig zijn geplaatst (er mag geen ruimte zijn tussen het kapje en de basis van de punt) en de plastic film geen gaten, scheuren of kreukels bevat. Als het kapje is geplaatst, plaatst u de sondepunt in het hoge temperatuurdoel, naast de knipperende LED. Controleer of de sondepunt volledig in doelopening is geplaatst. Stap 6: De plaatsing van de sondepunt in het hoge doel wordt automatisch opgemerkt en op het scherm wordt “Druk op scanknop op GENIUS 2-thermometer” weergegeven. Druk op de scanknop van de GENIUS 2-thermometer terwijl deze in het hoge temperatuurdoel is geplaatst. De GENIUS 2-thermometer geeft drie korte geluidssignalen, terwijl de temperatuur van het blackbody-doel wordt gemeten. Als de temperatuur is opgenomen wordt het volgende scherm weergegeven. Stap 7: Als de GENIUS 2-thermometer zich binnen bepaalde grenzen bevindt, wordt een resultatenscherm weergegeven met het serienummer van de thermometer en de bevestiging dat de thermometer is geslaagd voor de kalibratiecontrole. Als de GENIUS 2-thermometer zich buiten de opgegeven grenzen bevindt, wordt op de display aangegeven dat een kalibratieprocedure wordt uitgevoerd door de GENIUS 2-controle/kalibrator en dat de gebruiker stappen 3 tot en met 6 nog drie keer moet uitvoeren om herkalibratiegegevens te kunnen verzamelen. Nadat u stappen 3 tot en met 6 de laatste keer hebt uitgevoerd, voert de GENIUS 2-controle/kalibrator opnieuw een kalibratie uit van de GENIUS 2-thermometer. Als de herkalibratie is geslaagd, wordt een resultatenscherm weergegeven met het serienummer van de thermometer en de bevestiging dat de thermometer is geslaagd voor de kalibratie. Als de herkalibratie niet is geslaagd of als de thermometer niet opnieuw kan worden gekalibreerd, wordt een resultatenscherm weergegeven met het serienummer van de thermometer en het bericht dat de kalibratie van de thermometer is mislukt. Stap 8: Nadat het resultatenscherm is weergegeven, drukt u op de codeerknop om een tweede resultatenscherm weer te geven met de doeltemperaturen, de metingen van de GENIUS 2 en de afwijkingen tussen het doel en de thermometer. Stap 9: In het volgende scherm wordt “Rapport naar USB-flashstation schrijven?” weergegeven, met daaronder Ja en Nee. Als u een rapport wenst, plaatst u een Flash-station in de USB-poort op het voorpaneel. Let op: het duurt na het plaatsen ongeveer 30 seconden voordat het USB-flashstation is geïnitialiseerd. Gebruik de codeerknop om naar Ja of Nee te gaan en druk op de codeerknop om een keuze te maken. Als u Nee selecteert, keert u terug naar het hoofdmenu. Als u Ja selecteert, wordt een rapport geschreven naar het USB-flashstation, waarbij het serienummer van de thermometer wordt gebruikt als bestandsnaam en .TXT wordt gebruikt als bestandsextensie. Het bestand bevat de testresultaten en andere informatie, zoals serienummer, datum, tijd, gebruikersnaam en organisatie. Dit bestand is een eenvoudig ASCII-tekstbestand en is alleen beschikbaar in het Engels. Dit bestand kan worden gelezen door MicrosoftTM Windows-programma’s zoals Kladblok en kan naar een printer worden verzonden. Nadat het bestand is geschreven, wordt op het scherm “Druk op knop om verder te gaan” weergegeven. Druk op de codeerknop om terug te keren naar het hoofdmenu. Stap 10: Nadat een kalibratiecontrole/herkalibratie is voltooid, verwijderd u de sondeconnector uit het batterijgedeelte door de connector heen en weer te bewegen. Plaats vervolgens de drie AAA-batterijen terug en let hierbij op de juiste polariteit. Plaats ten slotte het deksel van het batterijgedeelte terug en schroef deze vast. GENIUS™ 2 Controle/kalibrator 8 10/23/2008 Artwork#s: WN 182563-01 Job#: 16245 Gebruiks- en onderhoudshandleiding Status: Approved Doc#: AW10001726 Effective Date: nulldate Stap 11: Verwijder het USB-flashstation voordat u het deksel van de behuizing sluit. Het deksel dient gesloten te zijn wanneer de GENIUS 2-controle/kalibrator niet in gebruik is Deel VII - Foutmeldingen Foutmeldingen zijn voorzien van een nummer, zoals fout 1, enzovoort. Hieronder volgt een uitleg van de fouten: Fout 1: Interne fout Deze fout geeft aan dat een probleem is opgetreden in de GENIUS 2-controle/kalibrator. Als deze fout zich blijft voordoen nadat u het apparaat hebt uitgeschakeld en opnieuw ingeschakeld, dient u de eenheid te retourneren voor onderhoud in de fabriek. Fout 2: Kan geen verbinding maken met GENIUS 2-thermometer. Deze fout geeft aan dat er geen communicatie of miscommunicatie plaatsvindt tussen de GENIUS 2-thermometer en de GENIUS 2-controle/kalibrator. Controleer beide uiteinden van de sondekabel van de thermometer, met name het uiteinde dat in het batterijgedeelte is aangesloten. De pinnen in de sondekabel van de thermometer verliezen mogelijk hun verende eigenschap wanneer ze zijn gebogen of beschadigd. Controleer de sondekabel en voer de kalibratiecontrole opnieuw uit. Probeer, indien mogelijk, een andere GENIUS 2-thermometer als het probleem zich blijft voordoen. Indien het probleem niet wordt opgelost, dient u de eenheid te retourneren voor onderhoud in de fabriek, samen met de thermometerkabel en de GENIUS 2-thermometer waarbij deze fout is opgetreden. Fout 3: Time-out Deze fout geeft aan dat de gebruiker te lang heeft gewacht met het uitvoeren van een handeling. Deze fout kan worden weergegeven bij het aansluiten van de GENIUS 2-thermometer of tijdens het installeren van een sondekapje op de thermometer. De time-out is ingesteld op één minuut. Voer de handeling waarbij deze fout is opgetreden opnieuw uit. Fout 4: Doel buiten bereik Deze fout geeft aan dat een van de blackbody-doelen niet binnen het bereik ligt van de temperatuur. Controleer of de eenheid wordt gebruikt binnen het bereik van de toegestane omgevingstemperatuur. Raadpleeg Deel VIII. Als de fout zich blijft voordoen wanneer u het apparaat gebruikt binnen de opgegeven omgevingstemperatuur, zijn de blackbody-doelen mogelijk beschadigd. Retourneer de eenheid voor onderhoud in de fabriek. Fout 5: Schrijffout van USB-flashstation Deze fout geeft aan dat een probleem is opgetreden bij het schrijven van het rapportbestand naar een USB-flashstation. Controleer of de schakelaar voor schrijfbeveiliging op het USB-station is uitgeschakeld. Controleer of er voldoende vrije ruimte op het station aanwezig is en of er geen beveiliging of codeerfuncties zijn ingeschakeld. Fout 6: Omgevingstemperatuur buiten bereik Deze fout geeft aan dat de omgevingstemperatuur niet binnen het opgegeven bereik ligt voor een nauwkeurige herkalibratie. Controleer of de eenheid wordt gebruikt binnen het bereik van de toegestane omgevingstemperatuur. Raadpleeg Deel VIII. Fout 7: Onjuist doel Deze fout geeft aan dat de GENIUS 2-thermometer in een onjuist doel is geplaatst of te vroeg in het doel is geplaatst. Volg de instructies op het scherm en voer geen stappen uit voordat de instructies worden weergegeven. De foutmelding wordt gedurende 60 seconden weergegeven voordat het hoofdmenu opnieuw wordt weergegeven. Druk op de codeerknop om de foutmelding eerder te sluiten. . GENIUS™ 2 Controle/kalibrator 9 10/23/2008 Artwork#s: WN 182563-01 Job#: 16245 Gebruiks- en onderhoudshandleiding Status: Approved Doc#: AW10001726 Effective Date: nulldate Deel VIII - Specificaties van controle/kalibrator Instelpunten van temperatuur Laag doel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32,2 °C +/- 0,3 °C (90,0 °F +/- 0,5 °F) Hoog doel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40,6 °C +/- 0,3 °C (105,0 °F +/- 0,5 °F) Nauwkeurigheid doeltemperatuur . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Initiële grens +/- 0,03 °C (0,05 °F) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Na 1 jaar +/- 0,06 °C (0,10 °F) Nauwkeurigheid GENIUS 2-thermometer na herkalibratie* Metingen tussen 36,0 °C en 39,0 °C (96,8 °F en 102,2 °F) . . . . . . . . . . . . . . . . . . . . . . . . . +/- 0,2 °C (+/- 0,4 °F) Metingen lager dan 36,0 °C (96,8 °F) of hoger dan 39,0 °C (102,2 °F) . . . . . . . . . . . . . . . . . +/- 0,3 °C (+/- 0,5 °F) Opwarmtijd voor gebruik . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15 minuten (vanuit een niet-extreme omgeving) Omgevingstemperatuur in bedrijf . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21,1 - 26,7 °C (70 - 80 °F) Relatieve vochtigheid in bedrijf . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50 +/- 20% zonder condensvorming Hoogte . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Tot 2000 meter (6563 feet) Opslagtemperatuur . . . . . . . . . . . . . . . . . . . . . . . . . . . -25 °C - 55 °C (-13 °F - 131 °F) 85% RL zonder condensvorming Gemiddelde afmetingen Lengte . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31,1 cm Diepte . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15,9 cm Breedte . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26,0 cm Gewicht . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3,3 kg Stroomvoorziening . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100-240 V 0,8-0,4 A 47-63 Hz Zekeringsvereisten . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Schurter FSF 5 x 20 mm snel smeltend . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Laag uitschakelvermogen 250 V, 3,15 A Aanbevolen kalibratie-interval van controle/kalibrator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Eén jaar ESD-immuniteit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Voldoet aan IEC 61000-4-2 Niet-continue Criteria C Mate van verontreiniging . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 Installatiecategorie. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . II . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Alleen voor gebruik binnenshuis *Nauwkeurigheid na herkalibratie met de GENIUS 2-controle/kalibrator is mogelijk niet gelijk aan fabriekskalibratie Deel IX - Reinigingsinstructies LET OP: DOMPEL DE CONTROLE/KALIBRATOR, DE KABELS OF HET NETSNOER NIET ONDER IN WATER OF REINIGINGSOPLOSSINGEN, REINIG DEZE MET EEN SCHONE, VOCHTIGE (DROGE) DOEK. ALS U ZICH NIET AAN DE IN DIT DOCUMENT BESCHREVEN REINIGINGSPROCEDURES HOUDT, KAN DIT GEVAAR VOOR DE GEBRUIKERS OPLEVEREN. Zoals geldt voor elk elektrisch apparaat, dient u erop te letten dat er geen vloeistof in de controle/kalibrator terechtkomt ter voorkoming van elektrische schokken, brand of schade aan elektrische componenten. GENIUS™ 2 Controle/kalibrator 10 10/23/2008 Artwork#s: WN 182563-01 Job#: 16245 Gebruiks- en onderhoudshandleiding Status: Approved Doc#: AW10001726 Effective Date: nulldate Algemene reinigingsinstructies Het reinigen van de GENIUS 2-controle/kalibrator kan als volgt worden uitgevoerd: • HAAL EERST DE STEKKER VAN DE GENIUS 2-CONTROLE/KALIBRATOR UIT HET STOPCONTACT. REINIG HET APPARAAT NOOIT ALS DIT NOG OP HET ELEKTRICITEITSNET IS AANGESLOTEN. • Voor een algemene reiniging van het buitenoppervlak dient een mild reinigingsmiddel te worden gebruikt. Een vochtige doek met isopropylalcohol kan worden gebruikt om de eenheid te reinigen. GEBRUIK GEEN schuurmiddelen of oplosmiddelen. Vermijd overmatige bevochtiging rond de doeleenheden en de elektrische connectoren. • Probeer niet het doel onder de rubberen flappen te reinigen. • Reinig de doelnesten of flappen niet met alcohol. Bij enige twijfel over de geschiktheid van een reinigingsmiddel voor een onderdeel van dit apparaat of het materiaal, neemt u contact op met de klantenservice van Tyco Healthcare/Kendall (zie Deel XI). Sectie X - Probleemoplossing Hieronder volgen enkele veelvoorkomende problemen: Probleem Mogelijke oorzaak Correcties GENIUS 2-kalibrator wordt niet ingeschakeld GENIUS 2-kalibrator is niet aangesloten op netstroom Controleer de netsnoeren Zekering doorgeslagen Vervang de zekering Testrapport wordt niet naar USBstation geschreven USB-flashstation is niet geformatteerd Gebruik een pc of laptop om het Flash-station te formatteren USB-flashstation is tegen schrijven beveiligd Gebruik een pc of laptop om schrijfbeveiliging uit te schakelen USB-flashstation beschikt niet over voldoende vrije ruimte Gebruik een pc of laptop om overbodige items te wissen of gebruik een ander USB-flashstation USB-flashstation is niet compatibel met GENIUS 2-controle/ kalibrator Gebruik een ander USB-flashstation GENIUS 2-kalibrator heeft onvoldoende tijd gehad om het USBflashstation te detecteren Probeer opnieuw te schrijven naar het USB-station Sondekapje zit vast in een van de blackbody-doelen Verwijder het sondekapje uit het blackbody-doel met een klein pincet of een hemostaat. Beschadig de binnenkant van het doel niet. Sondekapje is vuil Gebruik een nieuw sondeklepje voor elke meting. Gebruik een sondekapje niet opnieuw. Sondekapjes mogen niet met de hand worden aangeraakt. Thermometerlens is vuil Reinig de thermometerlens volgens de instructies in de gebruikshandleiding voor de GENIUS 2. GENIUS 2-thermometer is niet juist geplaatst Controleer of de GENIUS 2-thermometer volledig en onder de juiste hoek in het doel is geplaatst. GENIUS 2-thermometers, -sondekapjes en/of -controle/kalibrator is niet op kamertemperatuur. Controleer of de GENIUS 2-thermometers, -sondekapjes en/of -kalibrator voldoende tijd hebben gehad om op kamertemperatuur te komen voordat u verder gaat. Meerdere GENIUS 2-thermometers worden niet opnieuw gekalibreerd Fout 6 doet zich herhaaldelijk voor Mogelijk probleem met stroomverbruik met GENIUS 2-thermometer Herhaal de kalibratiecontrole/herkalibratie met een andere GENIUS 2-thermometer en retourneer de thermometer waarbij het probleem zich bij het testen van een bepaalde GENIUS 2-thermometer voordoet voor onderhoud in de fabriek. GENIUS 2-thermometer lijkt te hoog of te laag te meten na een geslaagde herkalibratie Sondekapje is opnieuw gebruikt na kalibratiecontrole/herkalibratie GENIUS™ 2 Controle/kalibrator Herhaal kalibratiecontrole/herkalibratie met een nieuw sondekapje voor elke plaatsing in het doel tijdens de procedure. 11 10/23/2008 Artwork#s: WN 182563-01 Job#: 16245 Gebruiks- en onderhoudshandleiding Status: Approved Doc#: AW10001726 Effective Date: nulldate Sectie XI - Klantenservice De GENIUS 2-controle/kalibrator moet eenmaal per jaar worden gekalibreerd en gecontroleerd in de fabriek. Neem contact op met Tyco Healthcare/Kendall en retourneer het apparaat op of voor de datum van kalibratie. Op de controle/kalibrator wordt één maand voor de datum van kalibratie een herinnering weergegeven. LET OP: HET APPARAAT BEVAT GEEN ONDERDELEN DIE ONDERHOUDEN KUNNEN WORDEN, VOOR ONDERHOUD RETOURNEREN NAAR DE FABRIEK. Indien het nodig is een eenheid voor reparatie te retourneren, gaat u als volgt te werk: 1. Neem contact op met Tyco Healthcare/Kendall zoals hieronder staat aangegeven of met uw vertegenwoordiger van de klantenservice voor het verkrijgen van een retourautorisatienummer en aanwijzingen voor verzending. 2. Verzend een verzekerd pakket naar uw regionale contactpersoon of de juiste onderstaande locatie. Verenigde Staten Tyco Healthcare Group LP 98.6 Faichney Drive Watertown, NY 13601, VS +1 800 448-0190 of (315) 788-5246 Onderdelenlijst Canada Tyco Healthcare Canada 7300 Trans Canada Highway Pointe-Claire, QC H9R 1C7 +1 877 664-8926 or (514) 695-1220 Buiten de VS en Canada Tyco Healthcare Group LP Unit 2 Talisman Business Center, London Road Bicester, OX26 6HR, VK +44-1869-328065 Neem contact op met uw plaatselijke klantenservice of uw verkoopvertegenwoordiger voor de onderstaande onderdelen. Beschrijving GENIUS 2-controle/kalibrator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .303097 Vernieuwde GENIUS 2-controle/kalibrator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .303096 Sectie XII - Garantie Tyco Healthcare/Kendall garandeert dat de GENIUS 2-controle/kalibrator vrij zal zijn van gebreken in materiaal en vakmanschap onder normaal gebruik gedurende een periode van één (1) jaar vanaf datum van levering door Tyco Healthcare/Kendall aan de eerste koper. De aansprakelijkheid is hierbij beperkt tot vervanging of reparatie, dit ter beoordeling van Tyco Healthcare/Kendall, door het servicecentrum van TycoHealthcare/Kendall van enig product dat onder normaal gebruik en service volgens Tyco Healthcare/ Kendall gebreken vertoont in materiaal of vakmanschap. Deze garantie is niet van toepassing op en Tyco Healthcare/Kendall zal niet verantwoordelijk zijn voor enig verlies geleden in verband met de aankoop van enig product dat is gerepareerd door enige andere partij dan Tyco Healthcare/Kendall of haar geautoriseerde vertegenwoordiger, of dat is blootgesteld aan onjuist gebruik, verwaarlozing of ongeluk, of waarvan het serienummer is gewijzigd, beschadigd of verwijderd, of dat is gebruikt op een andere wijze dan volgens de instructies verstrekt door Tyco Healthcare/Kendall. Tyco Healthcare/Kendall accepteert noch volmachtigt enige vertegenwoordiger of andere partij uit haar naam enige aansprakelijkheid te aanvaarden anders dan de aansprakelijkheid die hier uitdrukkelijk is vermeld. ALLE ANDERE AANSPRAKEN OP GARANTIE, EXPLICIET OF IMPLICIET, MET INBEGRIP VAN IMPLICIETE GARANTIES VAN VERKOOPBAARHEID OF GESCHIKTHEID VOOR EEN BEPAALD DOEL OF EEN BEPAALDE TOEPASSING ANDERS DAN HETGEEN UITDRUKKELIJK IS GESTELD IN DE BIJ HET PRODUCT BEHORENDE DOCUMENTATIE, WORDEN DOOR Tyco Healthcare/Kendall AFGEWEZEN. IN GEEN GEVAL ZAL Tyco Healthcare/Kendall AANSPRAKELIJK ZIJN VOOR ENIGE INCIDENTELE, INDIRECTE OF VERVOLGSCHADE IN VERBAND MET DE AANKOOP OF HET GEBRUIK VAN ZIJN PRODUCTEN GENIUS™ 2 Controle/kalibrator 12 10/23/2008 Artwork#s: WN 182563-01 Job#: 16245 Gebruiks- en onderhoudshandleiding Status: Approved Doc#: AW10001726 Effective Date: nulldate Sectie XIII - Verklaring elektromagnetische conformiteit De Genius 2-controle/kalibrator is vervaardigd en getest volgens de normen IEC/EN61010-1 en EN61326-1. De controle/kalibrator is bedoeld voor gebruik in de onderstaande gespecificeerde elektromagnetische omgeving. De gebruiker van de kalibrator dient ervoor te zorgen dat het apparaat in een dergelijke omgeving wordt gebruikt Emissietesten Conformiteit Geleid en uitgestraald Emissies (EN61326-1) Groep 1, Klasse B Harmonische stroom (EN61000-3-2) Klasse A Spanningsschommelingen en flikkering (EN61000-3-3) Wordt aan voldaan Immuniteit voor uitgestraalde storing (EN61000-4-3) Wordt aan voldaan Immuniteit voor geleide storing (EN61000-4-6) Wordt aan voldaan Immuniteit voor stroomfrequentie magnetisch veld (EN61000-4-8) Wordt aan voldaan Immuniteit voor spanningsdalingen en onderbrekingen (EN61000-4-11) Wordt aan voldaan Immuniteit voor elektrische snelle transiënt/bursts (EN61000-4-4) Wordt aan voldaan Immuniteit plotselinge dalingen (EN61000-4-5) Wordt aan voldaan Elektrostatische ontlading (EN61000-4-2)) Voldoet aan criteria C, Niet-continu GENIUS™ 2 Controle/kalibrator Richtlijn - elektromagnetische omgeving De kalibratoremissies zijn zeer laag en zullen naar verwachting geen nabije elektronische apparatuur storen. 13 10/23/2008 Artwork#s: WN 182563-01 Job#: 16245 Gebruiks- en onderhoudshandleiding
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98
  • Page 99 99
  • Page 100 100
  • Page 101 101
  • Page 102 102
  • Page 103 103
  • Page 104 104
  • Page 105 105
  • Page 106 106
  • Page 107 107
  • Page 108 108
  • Page 109 109
  • Page 110 110
  • Page 111 111
  • Page 112 112
  • Page 113 113
  • Page 114 114
  • Page 115 115
  • Page 116 116
  • Page 117 117
  • Page 118 118
  • Page 119 119
  • Page 120 120
  • Page 121 121
  • Page 122 122
  • Page 123 123
  • Page 124 124
  • Page 125 125
  • Page 126 126
  • Page 127 127
  • Page 128 128
  • Page 129 129
  • Page 130 130
  • Page 131 131
  • Page 132 132
  • Page 133 133
  • Page 134 134
  • Page 135 135
  • Page 136 136
  • Page 137 137
  • Page 138 138
  • Page 139 139
  • Page 140 140
  • Page 141 141
  • Page 142 142
  • Page 143 143
  • Page 144 144
  • Page 145 145
  • Page 146 146
  • Page 147 147
  • Page 148 148
  • Page 149 149
  • Page 150 150
  • Page 151 151
  • Page 152 152
  • Page 153 153
  • Page 154 154
  • Page 155 155
  • Page 156 156
  • Page 157 157
  • Page 158 158
  • Page 159 159
  • Page 160 160
  • Page 161 161
  • Page 162 162
  • Page 163 163
  • Page 164 164
  • Page 165 165
  • Page 166 166
  • Page 167 167
  • Page 168 168
  • Page 169 169
  • Page 170 170
  • Page 171 171
  • Page 172 172
  • Page 173 173
  • Page 174 174
  • Page 175 175
  • Page 176 176
  • Page 177 177
  • Page 178 178
  • Page 179 179
  • Page 180 180
  • Page 181 181
  • Page 182 182
  • Page 183 183
  • Page 184 184
  • Page 185 185
  • Page 186 186
  • Page 187 187
  • Page 188 188
  • Page 189 189
  • Page 190 190
  • Page 191 191
  • Page 192 192
  • Page 193 193
  • Page 194 194
  • Page 195 195
  • Page 196 196
  • Page 197 197
  • Page 198 198
  • Page 199 199
  • Page 200 200
  • Page 201 201
  • Page 202 202
  • Page 203 203
  • Page 204 204
  • Page 205 205
  • Page 206 206
  • Page 207 207
  • Page 208 208
  • Page 209 209
  • Page 210 210
  • Page 211 211
  • Page 212 212
  • Page 213 213
  • Page 214 214
  • Page 215 215
  • Page 216 216
  • Page 217 217
  • Page 218 218
  • Page 219 219
  • Page 220 220
  • Page 221 221
  • Page 222 222
  • Page 223 223
  • Page 224 224
  • Page 225 225
  • Page 226 226
  • Page 227 227
  • Page 228 228
  • Page 229 229
  • Page 230 230
  • Page 231 231
  • Page 232 232
  • Page 233 233
  • Page 234 234
  • Page 235 235
  • Page 236 236
  • Page 237 237
  • Page 238 238
  • Page 239 239
  • Page 240 240
  • Page 241 241
  • Page 242 242

Tyco Kendall Genius 2 Operation And Service Manual

Type
Operation And Service Manual